Great Britain’s New Medical Device Post-Market Surveillance Rules in Force Since June 2025
The Medical Devices (Post-market Surveillance Requirements) (Amendment) (Great Britain) Regulations 2024 took effect on 16 June 2025, introducing mandatory surveillance plans, shorter incident reporting deadlines and periodic safety reporting for devices placed on the Great Britain market. The rules apply to UKCA- and CE-marked devices, including software and AI as a medical device, and are the first major change to the GB device framework under the MHRA’s reform programme.
- Made 16 December 2024
- In force 16 June 2025
- PMS plan mandatory for all devices
- Serious incidents: 2, 10 or 15 days
- Covers UKCA and CE-marked devices
- Instrument
- The Medical Devices (Post-market Surveillance Requirements) (Amendment) (Great Britain) Regulations 2024, amending the Medical Devices Regulations 2002
- Issued by
- UK Government; enforced by the Medicines and Healthcare products Regulatory Agency (MHRA)
- Date
- Made 16 December 2024; in effect from 16 June 2025
- Applies to
- Manufacturers of medical devices, in vitro diagnostic devices and active implantable devices placed on the market or put into service in Great Britain, including UKCA- and CE-marked devices
- What you must do
- Maintain a documented post-market surveillance system and plan for every device or device family, report serious incidents within risk-based deadlines, report trends, and prepare periodic post-market surveillance or safety update reports according to risk class.
- Not covered
- Devices outside Great Britain. Sources indicate the new requirements do not apply to devices placed on the market or put into service before 16 June 2025.
- Primary source
- First major overhaul of medical device regulation comes into force across Great Britain
Medicines and Healthcare products Regulatory Agency, 16 June 2025 · read 9 November 2026
In force Great Britain
What the Regulations Cover
The Post-market Surveillance Regulations amend the Medical Devices Regulations 2002, which continue to govern devices placed on the market in Great Britain (England, Wales and Scotland). The amending instrument was made into law on 16 December 2024 and, after a six-month implementation period, came into effect on 16 June 2025. The MHRA described the change as the first major overhaul of medical device regulation across Great Britain.
The rules cover general medical devices, in vitro diagnostic devices and active implantable devices, whether they carry the UKCA mark or the CE mark. Standalone software and AI that qualify as medical devices are within scope in the same way as hardware. The MHRA enforces compliance, and non-compliance can lead to restrictions on placing products on the GB market.
The MHRA published guidance on the new requirements in January 2025 and updated it around the June start date. New GB data schemas for reporting went live in the Manufacturers Online Reporting Environment (MORE) on 16 June 2025.
What Changed for Manufacturers
Every manufacturer must now have a documented post-market surveillance system and a post-market surveillance plan for each device or device family, regardless of risk class. The requirements are proportionate to risk, and the plan is intended to drive active tracking of safety and performance once devices are in use.
Incident reporting deadlines have been shortened and tiered. Legal and consultancy summaries describe a maximum of 15 days for most serious incidents, 10 days for a death or an unanticipated serious deterioration in health, and 2 days for a serious public health threat. Sources differ on whether the 15-day period is counted in calendar or working days, so manufacturers should check the regulation text and MHRA guidance. Separate provisions require manufacturers to investigate serious incidents and submit a final report. One source notes that the period for reporting an anticipated serious deterioration in health was reduced from 30 to 15 days.
The scope of reportable events has been extended to include side effects, and trend reporting has been introduced: incidents that do not need to be reported individually must still be captured in the manufacturer’s surveillance data and reported to the MHRA under the trend reporting requirements. Manufacturers must also notify the MHRA of field safety corrective actions taken outside Great Britain where similar devices are on the GB market; one consultancy summary gives a deadline of three UK working days.
Periodic Reporting by Risk Class
Manufacturers must prepare either a post-market surveillance report or a periodic safety update report (PSUR), depending on the device’s risk classification, with timescales set by class. The MHRA has issued a separate PSUR guidance document for approved bodies that refers to regulation 44ZM. The specific reporting frequencies for each class were not confirmed from the sources checked, and manufacturers should take them from the MHRA guidance directly.
For software and AI products, these duties mean that monitoring of real-world performance, user complaints and software defects must be built into the surveillance plan from launch. The MHRA’s AI Airlock regulatory sandbox, which received £3.6 million in multi-year funding in 2026 according to trade reporting, has examined topics such as predetermined change control plans, but these remain under development and do not change the surveillance duties now in force.
Implications for Device Sponsors
Manufacturers that market the same device in Great Britain and the EU will find the GB requirements broadly comparable to EU post-market surveillance obligations, but the reporting routes, timelines and data schemas are specific to the MHRA. Sponsors relying on CE marking for GB access must still meet the GB post-market rules. UK Responsible Persons acting for overseas manufacturers should confirm that vigilance procedures route GB reports through MORE.
The regulations are the first stage of the MHRA’s wider reform of the GB framework; further changes to pre-market requirements are expected separately. Pubrica does not provide legal advice; manufacturers should rely on the regulations and the MHRA’s published guidance.
What to Do
Five steps for manufacturers supplying devices in Great Britain
Software manufacturers should include real-world performance monitoring and defect tracking in the PMS plan, since these feed directly into incident and trend reporting.
Who It Applies To
Subjects
All medical devices
General devices, IVDs and active implantables, including software and AI that qualify as medical devices.
Population
Manufacturers and UK Responsible Persons
Organisations placing UKCA- or CE-marked devices on the GB market, including overseas manufacturers.
Countries
Great Britain
The regulations apply in England, Wales and Scotland.
Frequently Asked Questions
They came into effect on 16 June 2025. The amending regulations were made into law on 16 December 2024.
Yes. Sources state that they apply to both UKCA- and CE-marked devices placed on the market in Great Britain.
Summaries of the regulations describe 2 days for a serious public health threat, 10 days for a death or unanticipated serious deterioration in health, and 15 days for other serious incidents. Sources differ on whether days are calendar or working days.
Yes. A documented post-market surveillance system and plan are required for every device or device family, regardless of risk class, with requirements proportionate to risk.
Sources
Each source was checked on the date shown. Requirements may change after that date; confirm the current version with the issuing body.
- 1Medicines and Healthcare products Regulatory Agency. (2025, June 16). First major overhaul of medical device regulation comes into force across Great Britain. GOV.UK press release.
https://www.gov.uk/government/news/first-major-overhaul-of-medical-device-regulation-comes-into-force-across-great-britain Read 9 November 2026. - 2Medicines and Healthcare products Regulatory Agency. (2025, January). MHRA guidance on new medical devices post-market surveillance requirements. GOV.UK press release.
https://www.gov.uk/government/news/mhra-guidance-on-new-medical-devices-post-market-surveillance-requirements Read 9 November 2026. - 3Medicines and Healthcare products Regulatory Agency. (2025). Medical devices: periodic safety update report. Guidance document.
https://assets.publishing.service.gov.uk/media/6852d79d679778c74ec15e7a/Medical_devices_-_periodic_safety_update_report.pdf Read 9 November 2026. - 4CMS. (2025). New post-market medical device surveillance regulations take effect in Great Britain. Law firm update.
https://cms.law/en/gbr/legal-updates/new-post-market-medical-device-surveillance-regulations-take-effect-in-great-britain Read 9 November 2026. - 5BSI. (2025, March). How to prepare for the new PMS requirements for Great Britain. Webinar slides.
https://www.bsigroup.com/siteassets/pdf/en/insights-and-media/media/webinars/bsi-md-prepare-for-pms-requirements-webinar-250325-en-gb.pdf Read 9 November 2026. - 6Fieldfisher. (2025). Great Britain’s new post-market surveillance requirements, six months on. Law firm article.
https://www.fieldfisher.com/en/insights/great-britain-s-new-post-market-surveillance-requi Read 9 November 2026. - 7Imperial College London. Idea to adoption part 9: post-market surveillance. University resource.
https://www.imperial.ac.uk/medtechone/medtechone-streams/collaborative/collaborative-resource-idea-to-adoption/idea-to-adoption-part-9--post-market-surveillance/ Read 9 November 2026. - 8Pharmafile. (2026). MHRA expands its pioneering AI Airlock programme with £3.6m funding boost. Trade news.
https://pharmafile.com/news/mhra-expands-its-pioneering-ai-airlock-programme-with-3-6m-funding-boost/ Read 9 November 2026.
MHRA post-market surveillanceUK medical device regulationsserious incident reportingPSURsoftware as a medical device UK
Next steps
Check How This Requirement Applies to Your Work
- 1Confirm whether the requirement applies to your product, study or submission
- 2Plan the documents and evidence needed before the deadline
- 3Keep a record of how each obligation was met